Sample Surface Cleaning
Routine Microvolume Sample Surface Cleaning
1. Pipette 2 μL of dH20 onto the bottom sample surface. Lower the upper arm. Wait 1-2 minutes.
2. Wipe away the water from both the upper and lower sample surfaces with a dry, lint-free lab wipe.
Between Measurements
Use a dry, lint-free lab wipe to remove samples from both the lower and upper sample surfaces.
Click here for more information.